CCM

Contact Current Meter

products

FREQUENCY RANGE 40 Hz - 110MHz

CONTACT CURRENT MEASUREMENT

CCM (Contact Current Meter) measures the contact current flowing through the human body, as the latter makes contact with a conductive object charged by an EM field.
This instrument can verify the compliance with the limits for the exposition to contact current for workers and general public, shown in the ICNIRP guidelines and compliant with the European Parliament directive 2013/35/EU.

The display indication provides the value of the current in mA and the percentage value in relation to the standard.

Chart for CCM tool Contact Current Meter
FREQUENCY RANGE REFERENCE LEVEL FOR WORKERS lc (mA) REFERENCE LEVEL FOR PUBLIC lc (mA)
0 - 2.5 kHz
1.00
0.50
2.5 - 100 kHz
0.40f
0.20f
0.1 - 110 MHz
40.00
20.00
FREQ. RANGE REF. LEVEL FOR WORKERS lc (mA) REF. LEVEL FOR PUBLIC lc (mA)
0 - 2.5 kHz
1.00
0.50
2.5 - 100 kHz
0.40f
0.20f
0.1 - 110 MHz
40.00
20.00

MEASUREMENT

HAND SETUP

GROUND PLANE SETUP

CCM Hand Ground

HAND SETUP

Contact Current Measurement: HAND Mode
CCM tool Contact Current Meter

The CCM measures the current flowing through the operator (human body impedance).

GROUND PLANE SETUP

Contact Current Measurement: GROUND PLANE mode
CCM tool Contact Current Meter

The CCM measures the current flowing through the ground plane (human body simulation impedance).

STANDARD CONFIGURATION
Rigid case
CCM
1.5 V batteries (2pcs)
Body simulation impedance (Z-2251
Ground plane plate (dimensions 360 X 239 mm)
Ground plane cover
USB cable for PC connection
USB key with:
  • User manual
  • Datasheet
Calibration certificat
OPTIONS
CCM-JIG kit including:
  • Calibration JIG
  • Standard resistance (R45)
  • Cable (length 1m) N(M)-sma(M)
IEC-60990 (50 mA) body simulation impedance
IEC-60990 (120 mA) body simulation impedance
CCM-TIC hand grab simulation

CCM JIG SETUP

This option enables to calibrate the CCM through signal generators in the frequency range 40 Hz – 110 MHz. To verify the limit over 40 mA at 10 MHz, a power amplifier is also required.

CCM tool calibration

MEASURE STORAGE

The CCM allows storing all measures, in an exportable file (CSV extension) including date, time, Workers value, General public value, both LF and HF ranges value and setup used.

TECHNICAL SPECIFICATIONS

FREQUENCY RANGE 40Hz ... 110 MHz
Low band
40 Hz ... 2,5 kHz
Medium band
2,5 kHz ... 100 kHz
High band
100 kHz ... 110 MHz
FREQUENCY RESPONSE
Low band (40 Hz ... 2,5 kHz) @ 1mA
< ± 1.5 dB
Medium band (2,5 kHz ... 100 kHz) @ 100%
< ± 1.5 dB
High band (100 kHz ... 100 MHz) @ 20mA
< ± 1.5 dB

MEASUREMENT RANGE

Low Band (40 Hz ... 2,5 kHz)
- Level range
0.01...3 mA (ICNIRP limit 1 mA)
- Damage level
100 mA
- Resolution
1 nA
- Dynamic range @ 500 Hz
50 dB
- Linearity error @ 500 Hz 0.3 ... 3mA
< ± 1 dB
Medium Band (2,5 kHz ... 100 kHz)
- Level range
1 ... 300 % (ICNIRP limit 1 to 40 mA) (Ty)
- Damage level
500%
- Resolution
1 nA
- Dynamic range @ 25 kHz
50 dB
- Linearity error @ 25 kHz 10 ... 200%
< ± 1 dB
High Band (100 kHz ... 110 MHz)
- Level range
0.4 ... 120 mA (ICNIRP limit 40 mA)
- Damage level
300 mA
- Resolution
10 nA
- Dynamic range @ 10 MHz
50 dB
- Linearity error @ 10 MHz 12 ... 120mA
< ± 1 dB
INPUT SIGNAL ATTENUATION
200 MHz → 7 dB
300 MHz → 18 dB
400 MHz → 31 dB
500 MHz ... 3 GHz → 45 dB
OTHER SPECIFICATIONS
Measurement modes
Hand and Ground Plane
Display
Graphic LCD with led backlight
Alarm sound
Programmable level
Detectors
RMS
Contact tip
Tip radius 2 mm interchangeable
USB Interface
Micro USB Connector
Standard
Directive 2013/35/EU
Operating temperature
From 10°C to 40°C
Power supply
  • Battery
  • Operation time

2 pcs Alkaline AA
48 h
Dimensions
205 x 90 x 45 mm
Weight
200g
Recommended calibration interval
24 months
Built-in self-test
Safety front-end functionality test

DOWNLOAD

Brochure of CCM tool Contact Current Meter
DATASHEET

GET IN TOUCH

CONTACTS